Hanna Instruments General Use Electrode Cleaning Solutions

Electrodes can become dirty from use and will produce inaccurate results even as they read correctly in a pH buffer. Hanna’s cleaning solutions eliminate impurities and residues that are left on electrode surfaces when immersed in samples during measurement or when stored incorrectly. Hanna suggests cleaning the bulb and junction of your electrode on a regular basis to ensure that the probe is always clean and to prevent any clogging of the junction.

  • Specially formulated for cleaning electrodes
  • Air-tight seal to ensure quality
  • Expiration date and lot number

The most common cause for pH measurement inaccuracies is an unclean or improperly cleaned electrode. This is very important to note, because during calibration, the instrument assumes that the electrode is clean, and that the standardization curve created during the calibration process will remain a valid reference until the next calibration. pH meters on the market today will allow an offset voltage of approximately ±60 mV. The deviation from 0 mV is not unusual but ideally should be no greater than ±30 mV. The calibration process compensates for the change in offset voltage. If the mV offset continues to deviate and the meter is calibrated with a dirty electrode, then inaccurate readings will result.

Part NumberApplicationSizePrice
HI7061L General Purpose500 ml$22.50
HI7073L Proteins500 ml$41.79
HI7074L Inorganic Substances500 ml$41.79
HI7077L Oils and Fats500 ml$41.79